FDA Adverse Event Malfunction Summary report: N

GYNECARE X-TRACT MORCELLATOR

MDR report key: 3030367 · Received April 2, 2013

Report

Report Number
2210968-2013-03254
Event Type
Malfunction
Date Received
April 2, 2013
Report Date
March 12, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K993801
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. UPON EVALUATION, THE FLEX COUPLER WAS BROKEN.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC SUPRACERVICAL HYSTERECTOMY. DURING SET UP FOR THE PROCEDURE, IT WAS DISCOVERD THAT THE DEVICE WOULD NOT DRIVE THE HANDPIECE. THE HANDPIECE COULD BE CONNECTED TO THE DEVICE BUT IT NEVER TURNED ON OR ROTATED. IT ALSO APPEARED THAT THE COUPLER WAS SPLIT IN HALF. A DIFFERENT DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
135265 GYNECARE X-TRACT MORCELLATOR LAPAROSCOPIC MORCELLATOR HET ETHICON, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1