FDA Adverse Event Malfunction Summary report: N

ADVANTAGE BALLOON CATHETER INFLATION DEVICE

MDR report key: 3030154 · Received April 2, 2013

Report

Report Number
2134265-2013-01876
Event Type
Malfunction
Date Received
April 2, 2013
Report Date
March 4, 2013
Manufacturer
BOSTON SCIENTIFIC - CORK
Product Code
MAV
PMA / PMN Number
K955869
Removal / Correction Number
90803893FA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MANUFACTURER - DEVICE NOT RETURNED THEREFORE ANALYSIS OF COMPLAINT DEVICE COULD NOT BE PERFORMED. THE BATCH NUMBER IS UNKNOWN AND THE MANUFACTURING RECORDS FOR THE COMPLAINT DEVICE COULD NOT BE REVIEWED. THE ROOT CAUSE FOR THIS COMPLAINT IS SUPPLIER DESIGN. INVESTIGATION OF THE ISSUE FOUND THAT THE VENDOR PROCESS FOR HUB TO CANNULA TRANSITION IS NOT OPTIMIZED TO PRODUCE THE INSERTION TOOL PART. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR AN ANGIOPLASTY PROCEDURE, THE WIRE COULD NOT BE INSERTED INTO THE TOOL. THE PT2 LS GUIDE WIRE COULD NOT BE INSERTED THROUGH THE ADVANTAGE INSERTION TOOL. THE PROCEDURE WAS COMPLETED WITH NON-BSC INSERTION TOOL. THERE WAS NO PATIENT COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134125 ADVANTAGE BALLOON CATHETER INFLATION DEVICE SYRINGE, BALLOON INFLATION MAV BOSTON SCIENTIFIC - CORK H74904527011

Patients

Seq Age Sex Outcome Treatment
1 GUIDEWIRE: PT2 LS