FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK SPIRIT

MDR report key: 3030113 · Received April 2, 2013

Report

Report Number
2183996-2013-00547
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 18, 2013
Report Date
May 31, 2013
Manufacturer
ROCHE DIABETES CARE AG
Product Code
LZG
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE SOFT COMPONENTS OF THE HOUSING ARE WORN DOWN HEAVILY AND RESTRICTED HANDLING OF THE PUMP IS A POSSIBLE IMPLICATION. THEREFORE MOISTURE MAY ENTER THE INSULIN PUMP AND DESTROY THE FUNCTIONALITY OF THE BUTTONS AND THE PUMP ELECTRONICS. THE PROBLEM MENTIONED BY THE CUSTOMER IS RELATED TO CARELESS HANDLING OF THE PRODUCT.

Description of Event or Problem · 1

ON (B)(6) 2013, IT WAS REPORTED THAT THE UP BUTTON ON THE PATIENT'S INFUSION DEVICE HAD FALLEN OF AND WAS NOT FUNCTIONING. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. THE INFUSION DEVICE WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134727 ACCU-CHEK SPIRIT INSULIN INFUSION DEVICE LZG ROCHE DIABETES CARE AG 00700006862 NA

Patients

Seq Age Sex Outcome Treatment
1 046 YR