FDA Adverse Event
Malfunction
Summary report: N
PROGRAMMING SOFTWARE
MDR report key: 3030111
·
Received April 2, 2013
Report
- Report Number
- 1644487-2013-00882
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- March 8, 2013
- Report Date
- March 8, 2013
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
.
Description of Event or Problem · 1
ON (B)(6) 2013, IT WAS REPORTED THAT THE LANGUAGE COULD NOT BE SET TO (B)(6). WHEN ATTEMPTING TO DO SO, THE USER COULD NOT HIT THE 'OK' BUTTON. A HARD RESET SUCCEEDED IN GETTING THE DEVICE TO WORK. ADDITIONAL INFORMATION WAS RECEIVED THAT TAKING OUT THE FLASHCARD AND PLACING IT BACK IN RESOLVED THE PROBLEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 133987 | PROGRAMMING SOFTWARE | PROGRAMMING SOFTWARE | LYJ | CYBERONICS, INC. | MODEL 250 | 1063524 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |