PROBE 8225101 5PK STD PRASS FLUSH TIP
Report
- Report Number
- 1045254-2013-00317
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 11, 2013
- Manufacturer
- MEDTRONIC XOMED, INC.
- Product Code
- ETN
- PMA / PMN Number
- K934426
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES. CONCOMITANT PRODUCT: PROBE 8225101, 5PK STD PRASS FLUSH TIP, LOT# 0206211978, MANUFACTURED OCTOBER 1, 2012, USE BEFORE SEPTEMBER 29, 2020. (B)(4). PRODUCT NOT RETURNED FOR ANALYSIS, NO INFORMATION AVALIABLE. METHOD: ACTUAL DEVICE NOT EVALUATED.
DEVICE RETURNED (B)(4) 2013. EVALUATION SUMMARY: PRODUCT ANALYSIS CONFIRMED THE COMPLAINT, FROM VISUAL EVALUATION, ONE PROBE ASSEMBLY EXHIBITED THE INNER CANNULA EXPOSED NEAR THE DISTAL END (CLOSED TO THE TIP OF THE DEVICE) SINCE THE COATING ON THE PROBE WAS TORN / DAMAGED LIKELY DUE TO CUSTOMER'S EXCESSIVE USE OF THE DEVICE. THE TORN COATING PIECE WAS NOT RETURNED FOR EVALUATION. THE SECOND PROBE EXHIBITED MINOR CIRCULAR RING AT PROXIMAL END WHICH INDICATES THAT THE PROBE WAS SNUGGED INTO THE HANDLE BY THE CUSTOMER. A SAMPLE HANDLE ASSEMBLY WAS TAKEN AND BOTH PROBES WERE FOUND TO SECURELY ATTACH INSIDE THE HANDLE ASSEMBLY WITHOUT ANY ISSUES. RESISTANCE READINGS WERE TAKEN AND WERE FOUND TO BE LESS THAN 2 OHMS MEETING THE REQUIRED MANUFACTURING SPECIFICATION. ADDITIONALLY, A FUNCTIONAL TEST WAS PERFORMED USING THE NIM NEURO 3.0 INTERFACE, AND NIM 3.0 STIMULATOR WHICH REVEALED THAT THE PROBE FUNCTIONED AS INTENDED / ABLE TO PROVIDE STIMULATION WITHOUT OBSERVING ANY ISSUES / ANOMALIES. METHOD: MICROSCOPIC INSPECTION. RESULTS: STRESS PROBLEM.
IT WAS REPORTED THAT THE PROBE WAS USED DURING A DEMO IN THE OR, THE SURGEON WAS NOT FAMILIAR WITH THE PROBE AND PUSHED TOO HARD WHEN STIMULATING. THE PROBE BENT AND WAS BENT BACK SEVERAL TIMES. THE SURGEON MENTIONED THAT PARTS OF THE PROBE WERE CRUMBLED INTO THE OPERATION FIELD. THE PARTS WERE TAKEN OUT OF THE OPERATION FIELD MANUALLY; THERE WAS NO REPORTED PATIENT IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 134879 | PROBE 8225101 5PK STD PRASS FLUSH TIP | STIMULATOR, NERVE | ETN | MEDTRONIC XOMED, INC. | 8225101 | 0206029248 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | PROBE 8225101 5PK STD PRASS FLUSH TIP |