ERGOMETER STORAGE CART
Report
- Report Number
- 1932056-2013-00001
- Event Type
- Other
- Date Received
- March 22, 2013
- Date of Event
- February 27, 2013
- Report Date
- March 21, 2013
- Manufacturer
- MEDICAL POSITIONING, INC.
- Product Code
- DPS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
MEDICAL POSITIONING INC, HAS BEEN SELLING THE STRESS ECHOBED WITH AN AVAILABLE STORAGE CART FOR OVER 15 YRS. THIS IS THE FIRST REPORT OF AN ERGOMETER CART TIPPING. MEDICAL POSITIONING, INC. HAS DETERMINED THAT THIS EVENT WAS CAUSED BY USER ERROR. HAD THE USER FOLLOWED THE INSTRUCTIONS IN THE OWNER'S MANUAL FOR LOCKING THE PRODUCT'S TRACK IN PLACE, THE EVENT WOULD NOT HAVE OCCURRED AT ALL.
WHILE THE EVENT WAS REPORTED TO MPI ON (B)(6) 2013, MEDICAL POSITIONING, INC. DID NOT BECOME AWARE OF THE HAZARD INVOLVED IN THE REPORT UNTIL (B)(6) 2013 BECAUSE THE CUSTOMER DID NOT RESPOND TO THE COMPANY'S CONTACT ATTEMPTS FOR ALMOST ONE (1) MONTH AFTER THE EVENT OCCURRED. THERE WERE NO DEATH(S) OR SERIOUS INJURIES AS A RESULT OF THE EVENT. HOWEVER, MEDICAL POSITIONING, INC HAS DETERMINED THAT A RECURRENCE OF THE EVENT COULD LEAD TO A SERIOUS INJURY SUCH AS A BONE BREAK OR SERIOUS CONTUSION. THE ERGOMETER (EXERCISE BIKE) WAS BEING STORED ON ITS STORAGE CART. IT WAS FOUND THAT AN ADJUSTMENT LOCK WAS LEFT UNLOCKED BY THE USER, ALLOWING THE ERGOMETER TO SLIDE TO ONE SIDE CAUSING THE SYSTEM TO BECOME UNBALANCED. THE ERGOMETER WAS BUMPED BY THE USER IN ITS UNBALANCED STATE CAUSING THE CART TO TIP. WHILE NO INJURY OCCURRED IN THIS INSTANCE, A RECURRENCE COULD CAUSE A BROKEN BONE IF A BODY PART WAS IN THE WAY OF THE DEVICE AS IT FELL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 120935 | ERGOMETER STORAGE CART | ERGOMETER STORAGE CART - ACCESSORY | DPS | MEDICAL POSITIONING, INC. | 11043 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |