FDA Adverse Event Injury Summary report: N

ASTODIA

MDR report key: 3026796 · Received March 25, 2013

Report

Report Number
3026796
Event Type
Injury
Date Received
March 25, 2013
Date of Event
March 21, 2013
Report Date
March 25, 2013
Manufacturer
STIHLER ELECTRONIC, GMBH
Product Code
HJN
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
AZ, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

ATTEMPTED TO START A PERIPHERAL IV ON AN INFANT USING THE STIHLER ELECTRONIC ASTODIA LIGHT TO VISUALIZE THE VEINS. LESS THAN ONE MINUTE AFTER TURNING ON THE LIGHT AND HAVING IT AGAINST THE INFANT'S SKIN THE LIGHT BECAME HOT. REMOVED IT AND THE LIGHT HAD LEFT WHAT APPEARED TO BE A FIRST DEGREE BURN APPROXIMATELY 1.5CM X 1CM. THE NEONATAL NP NOTIFIED AND A SKIN CONSULT INITIATED.FOLLOW-UP INFORMATION RECEIVED THAT THE INFANT HAD BLISTER AND OPEN SKIN. WOUND CONSULT SHOWED PARTIAL THICKNESS BURN TO BOTTOM OF FOOT. PER UNIT, DEVICE HAD TO BE KEPT IN A CERTAIN POSITION TO STAY ON. PER BIOMED, SOME INTERMITTENT SHORT IN THE CABLE, BUT IT WAS A UNIT BROUGHT IN FOR EVALUATION SO COULD NOT BE EXAMINED/REPAIRED BY BIOMEDICAL. CABLE WAS TAPED TO THE CASE AND UNIT WAS USED BY STAFF.WHAT WAS THE ORIGINAL INTENDED PROCEDURE?IV START.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
121834 ASTODIA TRANSILLUMINATOR HJN STIHLER ELECTRONIC, GMBH * *

Patients

Seq Age Sex Outcome Treatment
1 6 DAY Other| R