NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2013-00221
- Event Type
- Malfunction
- Date Received
- March 28, 2013
- Report Date
- March 1, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT RD900 NEOPUFF DEVICE WAS RECENTLY REPAIRED BY A TRAINED PERSONNEL AT FISHER & PAYKEL HEALTHCARE REGIONAL OFFICE IN (B)(4). THE FAULTY COMPONENTS ARE CURRENTLY IN TRANSIT TO FPH (B)(4) FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON RECEIPT OF THE COMPONENTS OF THE COMPLAINT DEVICE AND COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE COMPLAINT RD900 NEOPUFF DEVICE WAS REPAIRED BY TRAINED PERSONNEL AT THE FISHER & PAYKEL HEALTHCARE REGIONAL OFFICE IN (B)(4). THE VALVE ASSEMBLY OF THE COMPLAINT NEOPUFF DEVICE WAS RETURNED TO FPH (B)(4) FOR EVALUATION. RESULTS: VISUAL INSPECTION REVEALED THAT THE OUTLET PORT OF THE RETURNED VALVE ASSEMBLY WAS BROKEN. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 091102. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN ACCIDENTALLY DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE BROKEN OUTLET PORT OF THE VALVE ASSEMBLY IS MOST LIKELY TO HAVE BEEN CAUSED BY EXCESSIVE IMPACT DAMAGE. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF/PERIVENT INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT." THE COMPLAINT NEOPUFF DEVICE WAS FITTED WITH A NEW FRONT PANEL AND VALVE ASSEMBLY, GIVEN A PERFORMANCE CHECK, AND RETURNED TO THE CUSTOMER FULLY FUNCTIONING AS PER SPECIFICATIONS.
A DISTRIBUTOR IN (B)(6) REPORTED THAT THE GAS OUTLET PORT OF AN RD900 INFANT RESUSCITATOR WAS BROKEN. NO PATIENT CONSEQUENCE WAS REPORTED.
A DISTRIBUTOR IN (B)(6) REPORTED THAT THE GAS OUTLET PORT OF AN RD900 INFANT RESUSCITATOR WAS BROKEN. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 127975 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |