FDA Adverse Event
Malfunction
Summary report: N
I-STAT1 FUSO ANALYZER
MDR report key: 3026067
·
Received March 28, 2013
Report
- Report Number
- 2245578-2013-00051
- Event Type
- Malfunction
- Date Received
- March 28, 2013
- Date of Event
- July 5, 2010
- Report Date
- August 20, 2010
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- CGA
- PMA / PMN Number
- K001387
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). INCIDENT WAS IDENTIFIED FOLLOWING REVIEW OF THIRD PARTY MANUFACTURER GENERATED RECORDS AGAINST PROCEDURAL REQUIREMENTS.
Description of Event or Problem · 1
ON (B)(6) 2010, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT ANALYZER #101158 WOULD NOT ACTIVATE. THE ANALYZER WAS RETURNED ON (B)(6) 2010 AND DURING FAILURE ANALYSIS CAPACITOR C4 ON THE MAIN PCB WAS FOUND SHORTED. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT. APOC HAS DETERMINED THAT THE IDENTIFIED COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 128030 | I-STAT1 FUSO ANALYZER | I-STAT1 ANALYZER | CGA | ABBOTT POINT OF CARE | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |