FDA Adverse Event Malfunction Summary report: N

I-STAT1 FUSO ANALYZER

MDR report key: 3026067 · Received March 28, 2013

Report

Report Number
2245578-2013-00051
Event Type
Malfunction
Date Received
March 28, 2013
Date of Event
July 5, 2010
Report Date
August 20, 2010
Manufacturer
ABBOTT POINT OF CARE
Product Code
CGA
PMA / PMN Number
K001387
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INCIDENT WAS IDENTIFIED FOLLOWING REVIEW OF THIRD PARTY MANUFACTURER GENERATED RECORDS AGAINST PROCEDURAL REQUIREMENTS.

Description of Event or Problem · 1

ON (B)(6) 2010, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER WHO REPORTED THAT ANALYZER #101158 WOULD NOT ACTIVATE. THE ANALYZER WAS RETURNED ON (B)(6) 2010 AND DURING FAILURE ANALYSIS CAPACITOR C4 ON THE MAIN PCB WAS FOUND SHORTED. BASED ON THE INFORMATION AVAILABLE, THERE WERE NO PATIENT OR USER RELATED INJURIES ASSOCIATED WITH THIS COMPLAINT. APOC HAS DETERMINED THAT THE IDENTIFIED COMPONENT FAILURE WITHIN THE ANALYZER CIRCUITRY, MAY LEAD TO THE BATTERIES BECOMING UNCOMFORTABLY HOT TO TOUCH IN THE AREA OF THE BATTERY COMPARTMENT.