FDA Adverse Event Malfunction Summary report: N

ACCU-CHEK ® MOBILE TEST STRIPS

MDR report key: 3024371 · Received March 27, 2013

Report

Report Number
1823260-2013-01886
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 12, 2013
Report Date
March 27, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
LFR
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE EVENT OCCURRED IN (B)(6). WHILE THIS PRODUCT IS NOT SOLD IN THE UNITED STATES, IT IS LIKE OR SIMILAR TO A PRODUCT MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

REPORTER STATED THAT CUSTOMER RECEIVED THE FOLLOWING RESULTS ON THE MOBILE SYSTEM WITHIN 1 MINUTE: 3.0 MMOL/L AND 13.0 MMOL/L. DURING THE INVESTIGATION PERFORMED BY THE MANUFACTURER, THE RESULTS OBTAINED THAT DAY WERE 3.1 MMOL/L AND 16.8 MMOL/L WITHIN 2 MINUTES. NO ACTIONS TAKEN BASED ON DEVICE RESULTS. NO ADVERSE EVENT REPORTED. THE MANUFACTURER REQUESTED RETURN OF SUSPECT PRODUCT FOR EVALUATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
125354 ACCU-CHEK ® MOBILE TEST STRIPS BLOOD GLUCOSE MONITORING TEST STRIPS LFR ROCHE DIAGNOSTICS NA 278212

Patients

Seq Age Sex Outcome Treatment
1 058 YR