FDA Adverse Event Malfunction Summary report: N

ENDEAVOR RESOLUTE RX

MDR report key: 3024316 · Received March 27, 2013

Report

Report Number
9612164-2013-00342
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
January 22, 2013
Report Date
March 5, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
NIQ
PMA / PMN Number
P110013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION RESULTS: INHERENT RISK OF PROCEDURE - (DEFORMATION). PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (LESION MORPHOLOGY) - 95% STENOSIS AND SEVERE CALCIFICATION. (DEFORMATION PROBLEM). EVALUATION CONCLUSIONS: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (LESION MORPHOLOGY) - 95% STENOSIS AND SEVERE CALCIFICATION. INHERENT RISK OF PROCEDURE - (DEFORMATION). (B)(4).

Description of Event or Problem · 1

PHYSICIAN WAS ATTEMPTING TO DELIVER ONE ENDEAVOR RESOLUTE DRUG-ELUTING STENT TO A LESION IN THE RCA WITH 95% STENOSIS AND SEVERE CALCIFICATION BUT THE STENT COULD NOT PASS THE LESION. THE PHYSICIAN REMOVED THE DEVICE. EVALUATION SUMMARY: THE STENT WAS POSITIONED BETWEEN THE MARKER BANDS AS PER SPECIFICATIONS. THE 8TH AND 9TH PROXIMAL SEGMENTS WERE PINCHED. PLEASE NOTE THAT THIS DEVICE (ENDEAVOR RESOLUTE) IS DISTRIBUTED OUTSIDE THE UNITED STATES; HOWEVER, IT IS SIMILAR TO THE UNITED STATES DISTRIBUTED PRODUCT (RESOLUTE INTEGRITY).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
125913 ENDEAVOR RESOLUTE RX STENT, CORONARY, DRUG-ELUTING NIQ MEDTRONIC IRELAND 0006161138

Patients

Seq Age Sex Outcome Treatment
1 00062 YR