STOCKERT 70 RF GENERATOR
Report
- Report Number
- 9612355-2013-00012
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- February 27, 2013
- Report Date
- February 28, 2013
- Manufacturer
- STOCKERT GMBH
- Product Code
- DRF
- PMA / PMN Number
- P990071
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION STILL IN PROGRESS. A SUPPLEMENTAL REPORT OR DEVICE EVALUATION WILL BE SUBMITTED. CARTO 3 SYSTEM: MODEL #: M-4800-01, SERIAL #: (B)(4). NAVISTAR: MODEL #: UNKNOWN, LOT #: UNKNOWN. (B)(4).
EVALUATION SUMMARY. (B)(4). IT WAS REPORTED DURING AN ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) PROCEDURE, THE ABLATION WOULD RESET ON THE CARTO 3 SYSTEM SCREEN AND THE RECORDING SYSTEM INTERMITTENTLY DURING THE CASE. ALSO, THE TIP OF THE MAPPING CATHETER WOULD TURN GREEN. DURING THIS TIME, THE STOCKERT 70 SYSTEM WOULD CONTINUE ABLATING. IT WAS ADVISED TO REPLACE SOME OF THE CABLES IN ORDER TO SEE IF THE ISSUE WOULD RESOLVE. AFTER A FOLLOW UP, IT WAS STATED THAT THERE WERE NO STOCKERT 70 SYSTEM ISSUES DURING THE FOLLOWING CASE AND THE GENERATOR PERFORMED NORMALLY. THE BWI FIELD REPRESENTATIVE INDICATED THAT THE CUSTOMER DOES NOT REQUIRE A LOANER OR STOCKERT SERVICE AT THIS TIME. THE DEVICE HISTORY RECORD (DHR) FOR THE STOCKERT GENERATOR SERIAL NUMBER (B)(4) SHOWS THAT NO MANUFACTURING OR TEST FAILS WERE NOTED DURING THE MANUFACTURING CYCLE RELATED TO FUNCTIONALITY OF THE DEVICE. THE DEVICE MET ALL REQUIREMENTS PRIOR TO DISTRIBUTION.
IT WAS REPORTED DURING AN ATRIOVENTRICULAR NODAL REENTRANT TACHYCARDIA (AVNRT) PROCEDURE, THE ABLATION WOULD RESET ON THE CARTO 3 SYSTEM SCREEN AND THE RECORDING SYSTEM INTERMITTENTLY DURING THE CASE. ALSO, THE TIP OF THE MAPPING CATHETER WOULD TURN GREEN. DURING THIS TIME, THE STOCKERT 70 SYSTEM WOULD CONTINUE ABLATING. UPON REQUEST, ADDITIONAL INFORMATION ON THE EVENT WAS REQUESTED FROM THE BWI FIELD REPRESENTATIVE. THE CARTO 3 SYSTEM WAS RESETTING TO 0 AS IF THE ABLATION WAS COMPLETED EVEN THOUGH THE STOCKERT 70 SYSTEM WOULD CONTINUE ABLATING. THE PROCEDURE WAS ABLE TO BE CONTINUED AS THE PROBLEM WAS INTERMITTENT. IT WAS COMPLETED SUCCESSFULLY WITHIN 30 MINUTES AND THERE WAS NO PATIENT CONSEQUENCE. A 4MM NAVISTAR CATHETER UNDER TEMPERATURE CONTROL WAS IN USE. APPROXIMATELY 10-15 WATTS WAS BEING DELIVERED. THE NORMAL TEMPERATURE INCREASES AND IMPEDANCE DROPS WERE OBSERVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126249 | STOCKERT 70 RF GENERATOR | RADIOFREQUENCY GENERATOR AND ACCESSORIES/ACCESSORY CABLE | DRF | STOCKERT GMBH | M-5463-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |