PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3023984
·
Received March 27, 2013
Report
- Report Number
- 2032227-2013-01173
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- March 12, 2013
- Report Date
- March 12, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES.
Description of Event or Problem · 1
THE CUSTOMER REPORTED UNRESPONSIVE BUTTONS AND A FROZEN SCREEN. TROUBLESHOOTING WAS PERFORMED, BUT THE ISSUES WERE NOT RESOLVED. THE CUSTOMER STATED THAT NONE OF THE BUTTONS WERE WORKING, AND THE TIME ON THE SCREEN WAS NOT ADVANCING. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.