FDA Adverse Event Malfunction Summary report: N

SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

MDR report key: 3023953 · Received March 27, 2013

Report

Report Number
1416980-2013-07487
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
September 26, 2012
Report Date
March 9, 2013
Manufacturer
BAXTER HEALTHCARE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING EVALUATION OF A HOMECHOICE HARDWARE DEVICE THE ALARM WAS CONFIRMED THROUGH THE EVENT HISTORY LOG REVIEW . THIS ALARM IS INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS OF THE CASSETTE CANNOT BE COMPLETED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A SYSTEM ERROR 2240 ALARM (AIR IN LINE) WAS IDENTIFIED IN THE LOG, WHICH INDICATES A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE. THE ALARM OCCURRED ON (B)(6) 2012 AT 01:47:42. NO ADDITIONAL INFORMATION IS AVAILABLE.