TRANSEND EX 014/205 SOFT TIP
Report
- Report Number
- 3008853977-2013-00056
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- March 14, 2013
- Report Date
- March 14, 2013
- Manufacturer
- BOSTON SCIENTIFIC COSTA RICA
- Product Code
- DQX
- PMA / PMN Number
- K931584
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE HISTORY RECORD REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. MICROSCOPIC INSPECTION OF THE RETURNED GUIDEWIRE FOUND THAT THE GUIDEWIRE WAS SEPARATED AT 93.5 CM FROM ITS PROXIMAL END. THE GUIDEWIRE APPEARED TO BE BENT AT THE SEPARATED SITE. THE GUIDEWIRE WAS ALSO TWISTED ON THE DISTAL SEPARATED SITE. THE POLYTETRAFLUOROETHYLENE (PTFE) COATING WAS EXAMINED NO ANOMALIES WERE OBSERVED. THE DEVICE WAS SENT FOR ADDITIONAL ANALYSIS WHICH INCLUDED SEM (SCANNING ELECTRON MICROSCOPY). SEM ANALYSIS DETERMINED THAT THE FAILURE OCCURRED DUE TO A BENDING OVERLOAD FOLLOWED BY A FINAL TENSION OVERLOAD. BOTH SAMPLES PRESENTED FAILURES ON THE EXTERNAL WIRE WITH CORRESPONDING CHARACTERISTICS SUGGESTING MATCH BETWEEN THEM. NO MATCH ON THE INNER WIRE WAS POSSIBLE TO DETERMINE. BASED ON THE SEM ANALYSIS AND INVESTIGATION, THE FAILURE OCCURRED DUE TO A BENDING OVERLOAD FOLLOWED BY A FINAL TENSION OVERLOAD (BENT FIRST, AND THEN PUSHED OR PULLED). THE DEVICE DIRECTIONS FOR USE (DFU) WAS REVIEWED AND THE FOLLOWING WAS FOUND: "EXERCISE CARE IN HANDLING A GUIDEWIRE DURING A PROCEDURE TO REDUCE THE POSSIBILITY OF ACCIDENTAL BREAKAGE, BENDING OR KINKING." AS WELL, PER THE DEVICE DFU: "NEVER ADVANCE THE GUIDEWIRE AGAINST EXCESSIVE RESISTANCE WITHOUT FIRST DETERMINING THE REASON FOR RESISTANCE UNDER FLUOROSCOPY. EXCESSIVE FORCE AGAINST RESISTANCE MAY RESULT IN SEPARATION OF THE GUIDEWIRE TIP, DAMAGE TO THE CATHETER AND/OR VESSEL PERFORATION." BECAUSE THE ISSUE OCCURRED WHEN ATTEMPTING TO INSERT THE GUIDEWIRE INTO A MICROCATHETER DURING THE PROCEDURE, THE CAUSE FOR THE REPORTED AND OBSERVED ISSUE IS BELIEVED TO BE HANDLING DAMAGE DURING USE.
DURING PREPARATION, THE PHYSICIAN ATTEMPTED TO INSERT THIS GUIDEWIRE INTO A MICROCATHETER BUT THE GUIDEWIRE BROKE. THE BROKEN PART OF THE GUIDEWIRE WAS REMOVED FROM THE MICROCATHETER USING FORCEPS. THIS EVENT TOOK PLACE DURING PREPARATION AND OUTSIDE THE PATIENT. THERE WAS NO PATIENT INVOLVED.
DURING PREPARATION THE PHYSICIAN ATTEMPTED TO INSERT THIS GUIDEWIRE INTO A MICROCATHETER BUT THE GUIDEWIRE BROKE. THE BROKEN PART OF THE GUIDEWIRE WAS REMOVED FROM THE MICROCATHETER USING FORCEPS. THIS EVENT TOOK PLACE DURING PREPARATION AND OUTSIDE THE PATIENT. THERE WAS NO PATIENT INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 127106 | TRANSEND EX 014/205 SOFT TIP | WIRE, GUIDE, CATHETER | DQX | BOSTON SCIENTIFIC COSTA RICA | 15614444 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |