FDA Adverse Event Summary report: N

16" CENTURY STEAM STERILIZER

MDR report key: 3023836 · Received March 27, 2013

Report

Report Number
3005899764-2013-00028
Date Received
March 27, 2013
Date of Event
March 1, 2013
Report Date
March 27, 2013
Manufacturer
STERIS MEXICO, S. DE R.L. DE C.V.
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN INSPECTED THE STERILIZER, AND FOUND THE STEAM GENERATOR HEATER AND RELAY CONTACTOR TO BE DAMAGED. THE TECHNICIAN REPLACED THE COMPONENTS, TESTED THE STERILIZER, FOUND IT OPERATIONAL AND RETURNED IT TO SERVICE. THE UNIT WAS MANUFACTURED ON 10/7/2008, AND IS UNDER STERIS SERVICE CONTRACT. STERIS MONTERREY QUALITY REVIEWED THE DHR FOR THIS UNIT, AND FOUND IT TO BE MANUFACTURED ACCORDING TO SPECIFICATIONS. STERIS ENGINEERING CONFIRMED THAT THIS INCIDENT IS AN ISOLATED EVENT.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THE STEAM STERILIZER MADE A LOUD POPPING NOISE AND TRIPPED THE POWER. THE UNIT WAS SMOKING AND SHUT DOWN. THE FIRE DEPARTMENT WAS DISPATCHED TO THE FACILITY. NO EVACUATION OF THE FACILITY OCCURRED. NO INJURIES WERE REPORTED. NO PROCEDURAL DELAYS OR CANCELLATIONS WERE REPORTED. NO PROPERTY DAMAGE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
125088 16" CENTURY STEAM STERILIZER STERILIZER FLE STERIS MEXICO, S. DE R.L. DE C.V.

Patients

Seq Age Sex Outcome Treatment
1