CARTO® 3 SYSTEM
Report
- Report Number
- 3008203003-2013-00039
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- March 5, 2013
- Report Date
- March 5, 2013
- Manufacturer
- BIOSENSE WEBSTER, LTD (HAIFA, ISRAEL)
- Product Code
- DQK
- PMA / PMN Number
- K090017
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). IT WAS REPORTED THAT DURING AN AFIB PROCEDURE, THE CARTO 3 WAS NOT SHOWING A RED NAVISTAR TIP DURING RF. A THERMOCOOL SF NAVISTAR WAS BEING USED TO ABLATE AT 30 WATTS. THE STOCKERT APPEARED TO BE FUNCTIONING NORMALLY AND WAS SENDING ABLATION DATA TO THE RECORDING SYSTEM. THE NAVISTAR ICON TIP REMAINED GREEN ON CARTO 3. THERE WAS NO IMPEDANCE CHANGE OR ELECTROGRAM REDUCTION WAS SEEN. RF WAS BEING DELIVERED EVEN THOUGH THE GREEN TIP WAS NOT TURNING RED. PACING WAS SET TO NONE AND THE ABLATION CHANNEL ON THE RECORDING SYSTEM WAS DISABLED. NORMAL IMPEDANCES WERE DISPLAYED ON STOCKERT. BIOSENSE WEBSTER FIELD SERVICE ENGINEER REPLACED THE BACKPLANE BOARD AND SET UP A WET BATH. FIELD SERVICE ENGINEER RAN FULL SYSTEM TEST AND THE SYSTEM PASSED IT SUCCESSFULLY AND THE SYSTEM WAS READY TO USE. THE DEFECTIVE BACKPLANE WAS REPLACED AND WAS SENT TO CARTO MANUFACTURER (HTC) FOR FURTHER INVESTIGATION. THE CARTO MANUFACTURER FOUND THAT A PIECE OF SOLDER SHORTED TWO PINS OF THE CONNECTOR P2 AND THIS CAUSED THE REPORTED ISSUE. THE BACKPLANE WAS SENT TO SUBCONTRACTOR FOR FURTHER REPAIR. THE DEVICE HISTORY RECORD REVIEW WAS PERFORMED. NO ANOMALIES WERE NOTED IN MANUFACTURING OR SERVICE OF THIS DEVICE.
(B)(4).
IT WAS REPORTED THAT DURING AN AFIB PROCEDURE, THE CARTO 3 WAS NOT SHOWING A RED NAVISTAR TIP DURING RF. A THERMOCOOL SF NAVISTAR WAS BEING USED TO ABLATE AT 30 WATTS. THE STOCKERT APPEARED TO BE FUNCTIONING NORMALLY AND WAS SENDING ABLATION DATA TO THE RECORDING SYSTEM. THE NAVISTAR ICON TIP REMAINED GREEN ON CARTO 3. THERE WAS NO IMPEDANCE CHANGE OR ELECTROGRAM REDUCTION WAS SEEN. RF WAS BEING DELIVERED EVEN THOUGH THE GREEN TIP WAS NOT TURNING RED. PACING WAS SET TO NONE AND THE ABLATION CHANNEL ON THE RECORDING SYSTEM WAS DISABLED. NORMAL IMPEDANCES WERE DISPLAYED ON STOCKERT. CHANGING THE REDEL CABLE, REDEL ADAPTOR, NAVISTAR CABLE, NAVISTAR AND THE STOCKERT DID NOT RESOLVE THE ISSUE. THE CASE WAS COMPLETED BY CHANGING TO ANOTHER CARTO 3 AND THE NAVISTAR ICON TIP TURNED RED WITH ABLATION. NO PATIENT ADVERSE CONSEQUENCE IS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126488 | CARTO® 3 SYSTEM | COMPUTER, DIAGNOSTIC, PROGRAMMABLE | DQK | BIOSENSE WEBSTER, LTD (HAIFA, ISRAEL) | M-4800-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |