FDA Adverse Event Injury Summary report: N

SPRINT QUATTRO

MDR report key: 3023697 · Received March 27, 2013

Report

Report Number
2649622-2013-02578
Event Type
Injury
Date Received
March 27, 2013
Date of Event
February 23, 2013
Report Date
February 28, 2013
Manufacturer
MPRI
Product Code
LWS
PMA / PMN Number
P920015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US WHERE THE SAME MODEL IS DISTRIBUTED. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. (B)(4). CONCOMITANT PRODUCT: D284DRG IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD), IMPLANTED: (B)(6) 2011; 5076-58 IMPLANTABLE PACING LEAD IMPLANTED: (B)(6) 2011.

Additional Manufacturer Narrative · 1

PRODUCT EVENT SUMMARY: (B)(4) - THE PARTIAL LEAD WAS RETURNED IN SEGMENTS, WAS ANALYZED AND NO ANOMALIES WERE FOUND. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. PERFORMANCE DATA NOTED THAT A LEAD INTEGRITY ALERT WAS TRIGGERED FOR LEAD FAILURE PREDICTOR ON (B)(6) 2013. VENTRICULAR NON-SUSTAINED TACHYCARDIA LESS THAN OR EQUAL TO 210 MILLISECONDS (MS) OCCURRED BETWEEN (B)(6) 2013. VENTRICULAR FIBRILLATION LESS THAN 130MS AVERAGE VENTRICULAR CYCLE ON (B)(6) 2013. THERE WERE VENTRICULAR SHORT INTERVAL COUNTS EQUAL TO 4805 BETWEEN (B)(6) 2013. THE DAILY PACE LEAD TREND DATA SHOWS AN INCREASE FOR VENTRICULAR PACING BIPOLAR IMPEDANCE FROM 532 OHMS TO 4047 OHMS PEAK BETWEEN (B)(6) 2013.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT RECEIVED EIGHT SHOCKS AND THAT A LEAD INTEGRITY ALERT. IT WAS FURTHER REPORTED THAT THERE WAS HIGH IMPEDANCE, OVERSENSING, NOISE AND A FRACTURE WAS SUSPECTED OF THE RIGHT VENTRICULAR LEAD. DURING EXTRACTION OF THE PACING AND TACHY LEADS THE PATIENT HAD A DECREASE IN VITAL SIGNS AND WAS TAKEN URGENTLY TO SURGERY. THE LEADS WERE REMOVED AND REPLACED AND THE PATIENT WAS TAKEN TO INTENSIVE CARE. THE CURRENT STATUS OF THE PATIENT HAS BEEN REQUESTED AND NOT RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT RECEIVED EIGHT SHOCKS AND THAT A LEAD INTEGRITY ALERT. IT WAS FURTHER REPORTED THAT THERE WAS HIGH IMPEDANCE, OVERSENSING, NOISE AND A FRACTURE WAS SUSPECTED OF THE RIGHT VENTRICULAR LEAD. DURING EXTRACTION OF THE PACING AND TACHY LEADS THE PATIENT HAD A DECREASE IN VITAL SIGNS AND WAS TAKEN URGENTLY TO SURGERY. THE LEADS WERE REMOVED AND REPLACED AND THE PATIENT WAS TAKEN TO INTENSIVE CARE. THE CURRENT STATUS OF THE PATIENT HAS BEEN REQUESTED AND NOT RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126973 SPRINT QUATTRO DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER LWS MPRI 694758

Patients

Seq Age Sex Outcome Treatment
1 00012 YR Hospitalization| L| R