MEGASUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2013-00972
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- February 18, 2013
- Report Date
- February 25, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY WAS ABLE TO SPIN FREELY AND IT WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT WRIST. OTHER CABLES AT WRIST WERE NOT DAMAGED. ADDITIONAL FINDING NOT REPORTED BY THE SITE WAS SURFACE SCRATCHES ON THE DISTAL PULLEY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT DURING A DA VINCI SI SACROCOLPOPEXY PROCEDURE, WHILE FINISHING SUTURING MESH, THE CUSTOMER NOTICED A WIRE PROTRUDING FROM DISTAL TIP OF THE MEGA SUTURE CUT NEEDLE DRIVER INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.