FDA Adverse Event
Malfunction
Summary report: N
IN TOUCH EU
MDR report key: 3023575
·
Received March 27, 2013
Report
- Report Number
- 0001831750-2013-02598
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- March 4, 2013
- Report Date
- March 4, 2013
- Manufacturer
- STRYKER MEDICAL-KALAMAZOO
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
SUPPLEMENTAL SUBMITTED AS INVESTIGATION FOUND THAT ONLY ELECTRONIC ENGAGEMENT OF THE BRAKES WAS AFFECTED. THE BRAKES COULD STILL BE ENGAGED MANUALLY. THEREFORE, THIS ISSUE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO SERIOUS INJURY OR DEATH IF IT WERE TO RECUR.
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE BRAKES WERE NOT ENGAGING DUE TO MALFUNCTION HEAD AND FOOT END CASTERS. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Description of Event or Problem · 1
IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE BRAKES WERE NOT ENGAGING ELECTRONICALLY DUE TO MALFUNCTIONED HEAD AND FOOT END CASTERS. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 125746 | IN TOUCH EU | BED, AC-POWERED ADJUSTABLE HOSPITAL | FNL | STRYKER MEDICAL-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |