FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 3023465 · Received March 27, 2013

Report

Report Number
1416980-2013-07434
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 1, 2013
Report Date
March 6, 2013
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
PMA / PMN Number
K071222
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS CONFIRMED. VISUAL EXAMINATION OF THE SAMPLE SHOWED LEAKAGE FROM HALF-BROKEN TUBING AT THE JUNCTION OF THE LUER POST. THE ROOT CAUSE WAS UNDETERMINED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT OF AN INFUSOR THAT LEAKED. THE CUSTOMER STATED THE LEAK WAS NOTED UPON UNPACKING THE PRODUCT. NO SOLUTION WAS LEFT IN THE DEVICE; HOWEVER, THE SOLUTION WAS CONTAINED IN THE OVERPOUCH. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.