FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3023424 · Received March 27, 2013

Report

Report Number
2531779-2013-03136
Event Type
Malfunction
Date Received
March 27, 2013
Report Date
March 19, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: REBOOTS ARE OBSERVED IN THE BLACK BOX. THERE WAS NO DAMAGE FOUND TO THE BATTERY COMPARTMENT. THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. A TEST CAP WAS USED FOR ALL INVESTIGATION STEPS. PUMP POWERS ON NORMAL WITH NO ALARMS. THE PUMP WAS EXERCISED FOR 24 HOURS WITH NO ALARMS OR POWER ISSUES OCCURRING. THE PUMP WAS OPENED AND NO DAMAGE WAS FOUND TO THE POWER CIRCUIT. DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS ON (B)(6) 2013 REPORTING THAT AFTER A BATTERY CHANGE, THE PUMP KEPT GOING TO THE VERIFY SCREEN. THE PATIENT REPORTED CHANGING THE BATTERY THREE TIMES AND THE ISSUE WAS RECURRING. THE PATIENT CONFIRMED NO YELLOW O-RING VISIBLE AND CONFIRMED NO KNOWN TRAUMA TO THE PUMP. THERE WAS NO REPORTED ADVERSE EVENT RELATED TO THIS ISSUE. THIS REPORT IS MADE BASED ON THE ALLEGATION OF THE INTERMITTENT POWER ISSUE WHICH WAS NOT RESOLVED WITH TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126283 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 48 YR