FDA Adverse Event Malfunction Summary report: N

ENTERALITE INFINITY ENTERAL FEEDING PUMP

MDR report key: 3023283 · Received February 28, 2013

Report

Report Number
1722139-2013-00136
Event Type
Malfunction
Date Received
February 28, 2013
Date of Event
December 1, 2012
Report Date
February 19, 2013
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
LZH
PMA / PMN Number
K031199
Removal / Correction Number
Z-0294-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT THE DIODE D7 WAS DETERMINED TO BE LEAKY AND WAS OUT OF THE SPECIFICATION. NEW PCB BOARD WAS REPLACED TO SOLVE THE ISSUE. REFERENCE RECALL # Z-0294-2013.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THAT PUMP ALARMS ERROR CODE 13.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
88963 ENTERALITE INFINITY ENTERAL FEEDING PUMP LZH MOOG MEDICAL DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1