FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 3023187
·
Received February 27, 2013
Report
- Report Number
- 2518422-2013-00328
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- February 4, 2013
- Report Date
- February 4, 2013
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVAL OF THE DEVICE AT THE MFR'S SVC CENTER, THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
A VENTILATOR WAS RETURNED TO THE MFR FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 84990 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC. | 1054655 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |