ZIMMER AIR DERMATOME HANDPIECE
Report
- Report Number
- 1526350-2013-00097
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- January 1, 2013
- Report Date
- January 30, 2013
- Manufacturer
- ZIMMER SURGICAL
- Product Code
- GFD
- Removal / Correction Number
- RES 61798
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
BEGINNING (B)(36) 2012, US AND CANADIAN CUSTOMERS WERE SENT AN URGENT PATIENT SAFETY ADVISORY INFORMING THEM OF THE NEED FOR PROPER CARE AND PREVENTIVE MAINTENANCE OF THEIR ZIMMER AIR DERMATOME. CUSTOMERS WERE INFORMED THAT IMPROPERLY MAINTAINED INSTRUMENTS MAY CAUSE DONOR SITE INJURIES OR RESULT IN DAMAGE TO THE GRAFT. CUSTOMERS WERE REQUESTED TO CONTACT ZIMMER TO SCHEDULE MAINTENANCE FOR THE DEVICE IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR REPAIR AND EVALUATION. THE SERVICE RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED ON 08/20/2007 AND HAS NO PREVIOUS REPAIR HISTORY. THE INSPECTION FOUND DAMAGE TO THE HEAD AND CONTROL BAR OF THE DEVICE. PRIOR TO REPAIR, THE DEVICE WAS OUTSIDE OF CALIBRATION SPECIFICATIONS AT ALL THICKNESS SETTINGS. DURING REPAIR, IT WAS NOTED THAT THE HOSE RETURNED WITH THIS UNIT WAS LEAKING. DURING CLINICAL FOLLOW UP, THE CUSTOMER STATED THE DEVICE WAS UTILIZED AT 150 PSI WITH NO EXTENSION CODE. PER INSTRUCTIONS FOR USE, "RECOMMENDED PRESSURE IS 100 PSI RUNNING, USING SUPPLIED HOSE. THE CAUSE IS MOST LIKELY DUE TO THE USER NOT MAINTAINING THE DEVICE PER PREVENTIVE MAINTENANCE AND HANDLING ACCORDING TO THE INSTRUCTIONS FOR USE. THE ZIMMER AIR DERMATOME SHOULD BE RETURNED EVERY 12 MONTHS AND THE HOSE EVERY 6 MONTHS FOR INSPECTION AND PREVENTIVE MAINTENANCE. ANNUAL FACTORY CALIBRATION CHECKS ARE STRONGLY RECOMMENDED TO VERIFY CONTINUED ACCURACY. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.
IT WAS REPORTED THAT THE ZIMMER AIR DERMATOME WAS CUTTING UNEVEN. ADDITIONAL CLINICAL FOLLOW UP WITH THE HOSPITAL INDICATED THAT THE DEVICE WAS UTILIZED AT 150 PSI WITH NO EXTENSION CODE. THERE WAS NO REPORT OF HARM, INJURY, DELAY, INCREASED SURGICAL TIME, OR MEDICAL/SURGICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 84992 | ZIMMER AIR DERMATOME HANDPIECE | ZIMMER AIR DERMATOME HANDPIECE | GFD | ZIMMER SURGICAL | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |