AIR DERMATOME HANDPIECE
Report
- Report Number
- 1526350-2013-00093
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- January 1, 2013
- Report Date
- December 17, 2012
- Manufacturer
- ZIMMER SURGICAL
- Product Code
- GFD
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- UNKNOWN
Narratives
THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR REPAIR AND EVALUATION. THE SERVICE RECORD INDICATES THAT THE DEVICE IS 12 YEARS OLD AND WAS LAST RETURNED TO THE MANUFACTURER FOR REPAIR ON (B)(4) 2007. INVESTIGATION OF THE DEVICE DISPLAYED DAMAGE TO THE HEAD AND CONTROL BAR. TESTING OF THE HOSE REVEALED THAT IT WAS DAMAGED. THE INITIAL INSPECTION ALSO NOTED THAT THE SWIVEL ON THE UNIT SEEMED VERY LOOSE. PRIOR TO REPAIR, THE DEVICE WAS OUTSIDE OF CALIBRATION AT THE ZERO THICKNESS SETTING ON THE LEFT AND RIGHT SIDE. THE DEVICE WAS ALSO OUTSIDE OF CALIBRATION ON THE LEFT SIDE AT THE 0.01 AND 0.02 THICKNESS SETTING. SIDE TO SIDE CALIBRATION WAS OUT AT THE 0, 0.01, 0.02 AND 0.03 THICKNESS SETTINGS. THE CAUSE IS MOST LIKELY DUE TO THE USER NOT MAINTAINING THE DEVICE PER PREVENTIVE MAINTENANCE AND HANDLING ACCORDING TO THE INSTRUCTIONS FOR USE. THE ZIMMER AIR DERMATOME SHOULD BE RETURNED EVERY 12 MONTHS AND THE HOSE EVERY 6 MONTHS FOR INSPECTION AND PREVENTIVE MAINTENANCE. ANNUAL FACTORY CALIBRATION CHECKS ARE STRONGLY RECOMMENDED TO VERIFY CONTINUED ACCURACY. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.
IT WAS REPORTED THAT AIR LEAKED FROM THE HOSE COUPLING AND SWIVEL COUPLING. IT WAS ALSO REPORTED THAT THE ZIMMER DERMATOME ALSO WORKED INTERMITTENTLY. ADDITIONAL CLINICAL FOLLOW UP WITH THE HOSPITAL INDICATED THAT THE REPORTED EVENT OCCURRED DURING STERILE PROCESSING AND THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 85119 | AIR DERMATOME HANDPIECE | AIR DERMATOME HANDPIECE | GFD | ZIMMER SURGICAL | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |