FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3023167
·
Received February 27, 2013
Report
- Report Number
- 8020893-2013-00478
- Event Type
- Malfunction
- Date Received
- February 27, 2013
- Date of Event
- January 25, 2013
- Report Date
- February 5, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SVC REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING. THERE WAS NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE ANALOG INTERFACE PCB. THE UNIT PASSED EXTENDED SELF-TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 85117 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |