FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3023132 · Received February 27, 2013

Report

Report Number
8020893-2013-00458
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 1, 2013
Report Date
February 1, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR...
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR HAD A BLANK SCREEN. THERE WAS NO PT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) CPU PCB AND BACKLIGHT INVERTER PCBS. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
84531 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR... 840

Patients

Seq Age Sex Outcome Treatment
1