FDA Adverse Event Malfunction Summary report: N

ACUITY CENTRAL MONITORING

MDR report key: 3023102 · Received February 27, 2013

Report

Report Number
3023750-2013-00026
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 1, 2013
Report Date
February 1, 2013
Manufacturer
WELCH ALLYN PROTOCOL, INC.
Product Code
DSI
PMA / PMN Number
K052160
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE BMET CONFIRMED THA THE CPU SPEAKER FAILED. THE CPU WILL NOT BE RETURNED FOR EVALUATION. THE ACUITY CPU SPEAKER IS AN OFF-THE-SHELF COMPUTER SUB-COMPONENT MADE BY ANOTHER MFR AND SOLD BY WELCH ALLYN. WELCH ALLYN DOES NOT POSSES TROUBLESHOOTING CAPABILITY BEYOND IDENTIFYING THESE SUB-COMPONENTS AS THE SOURCE OF THE FAILURE. METHOD: BMET CONFIRMED FAILURE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INTERNAL SPEAKER IN THEIR ACUITY SYSTEM IS NOT WORKING. THE BMET ATTACHED AN EXTERNAL SPEAKER FOR AUDIBLE ALARMS. THERE WAS NO REPORT OF ANY PT HARM AS A RESULT OF THE REPORTED EVENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85102 ACUITY CENTRAL MONITORING DSI WELCH ALLYN PROTOCOL, INC. ACUITY 6.40.01

Patients

Seq Age Sex Outcome Treatment
1