FDA Adverse Event Malfunction Summary report: N

ACUITY CENTRAL MONITORING

MDR report key: 3023074 · Received February 27, 2013

Report

Report Number
3023750-2013-00023
Event Type
Malfunction
Date Received
February 27, 2013
Date of Event
February 6, 2013
Report Date
February 6, 2013
Manufacturer
WELCH ALLYN PROTOCOL, INC.
Product Code
DSI
PMA / PMN Number
K022453
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE BMET CONFIRMED THAT THE CPU COOLING FAN FAILED. WELCH ALLYN TECH SERVICE ASSISTED THE BMET IN TROUBLESHOOTING THE PROBLEM. THE BMET VERIFIED THAT THE CPU FAN IS MAKING LOUD NOISES AND NOT SPINNING WELL. THE SYSTEM IS AN OLD SYSTEM THAT IS NO LONGER REPAIRABLE AT WELCH ALLYN. THE BMET SAID HE WOULD LOOK INTO WHAT HE COULD DO TO REPAIR THE FAN AT THEIR SITE. THIS SYSTEM IS SUBJECT OF AN END OF LIFE LETTER TO CUSTOMER AND IS NO LONGER REPAIRABLE AT WELCH ALLYN. THE ACUITY COOLING FAN IS AN OFF-THE-SHELF COMPUTER SUB-COMPONENT MADE BY ANOTHER MANUFACTURER AND SOLD BY WELCH ALLYN. WELCH ALLYN DOES NOT POSSESS TROUBLESHOOTING CAPABILITY BEYOND IDENTIFYING THESE SUB-COMPONENTS AS THE SOURCE OF THE FAILURE. METHOD: (BMET CONFIRMED FAILURE).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEIR ACUITY SYSTEM LOCKED UP AND HAD A SCRAMBLED SCREEN. UPON SUBSEQUENT REBOOT HE GETS THE ERROR "CPU FAN NOT STARTING. - HOT." THIS RESULTED IN A TEMPORARY INABILITY TO CENTRALLY MONITOR PATIENTS, HOWEVER BEDSIDE MONITORING WAS NOT AFFECTED. THERE WAS NO REPORT OF ANY PATIENT HARM AS A RESULT OF THE REPORTED EVENTS. THE CUSTOMER DID NOT PROVIDE ANY PATIENT INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85006 ACUITY CENTRAL MONITORING DSI WELCH ALLYN PROTOCOL, INC. ACUITY 6.31.00

Patients

Seq Age Sex Outcome Treatment
1