FDA Adverse Event
Malfunction
Summary report: N
T4 POWER PACK
MDR report key: 3023058
·
Received March 27, 2013
Report
- Report Number
- 0001811755-2013-00615
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- March 6, 2013
- Report Date
- March 6, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- LYU
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DURING THE DEVICE EVALUATION, THE REPORTED THERMAL DAMAGE TO THE PRONG COULD NOT BE CONFIRMED. IT WAS FOUND THAT THE DAMAGE TO THE BATTERY WAS NOT THERMAL DAMAGE. THE DAMAGE MAY HAVE BEEN CAUSED BY FORCE OR IMPACT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING CHARGING AT THE USER FACILITY, THE PRONG OF THE T4 POWER PACK LOOKED MELTED. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 125956 | T4 POWER PACK | ACCESSORY, SURGICAL APPAREL | LYU | STRYKER INSTRUMENTS-KALAMAZOO | 11341 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |