FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3023044 · Received March 27, 2013

Report

Report Number
3004209178-2013-92150
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 23, 2013
Report Date
March 25, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S INSULIN PUMP WAS NOT DELIVERING INSULIN. IT WAS STATED THAT THE RESERVOIR AND INFUSION SET WERE CHANGED TO RUN THE MANUAL PRIME TEST AND THE INSULIN DID NOT EXIT. THE HIGH PRESSURE TEST WAS PERFORMED TWICE AND THE DEVICE DID NOT ALARM NO DELIVERY. NO FURTHER INFORMATION WAS PROVIDED.