FDA Adverse Event Malfunction Summary report: N

TRILOGY 200

MDR report key: 3023025 · Received March 21, 2013

Report

Report Number
2518422-2013-00425
Event Type
Malfunction
Date Received
March 21, 2013
Date of Event
February 23, 2011
Report Date
February 23, 2011
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S AC INLET CONNECTOR WAS REPLACED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTIONS THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR PART 803 AND THE MANUFACTURER'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR'S AC INLET CONNECTOR WAS DAMAGED. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
117699 TRILOGY 200 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1032804

Patients

Seq Age Sex Outcome Treatment
1