FDA Adverse Event
Malfunction
Summary report: N
TPS UNIVERSAL DRIVER
MDR report key: 3023021
·
Received March 27, 2013
Report
- Report Number
- 0001811755-2013-00612
- Event Type
- Malfunction
- Date Received
- March 27, 2013
- Date of Event
- March 6, 2013
- Report Date
- March 6, 2013
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
UPON DISASSEMBLY FOR VISUAL INSPECTION, IT WAS FOUND THAT THE HOUSING WAS BROKEN.
Description of Event or Problem · 1
IT WAS REPORTED DURING SERVICE AT THE MANUFACTURER THAT THE TPS UNIVERSAL DRIVER CONTINUED TO RUN WHEN THE TRIGGER WAS NO LONGER ACTIVATED. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 126666 | TPS UNIVERSAL DRIVER | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |