FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3023012 · Received March 27, 2013

Report

Report Number
3004209178-2013-92126
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 25, 2013
Report Date
March 25, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARM DURING BASIC OCCLUSION TEST DUE TO LOOSE PROTRUDED DRIVE SUPPORT DISK. UNIT WAS RECEIVED WITH EXCESSIVE ADHESIVE AROUND THE END CAP. CRACKED END CAP, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND WINDOW NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE END CAP FELL OFF, AND THE INSULIN PUMP ALARMED. THE BLOOD GLUCOSE READING WAS 7.1MMOL/L. IT WAS STATED THAT THE DEVICE WAS BUMPED AND CRACKED. IT WAS ALSO STATED THAT CUSTOMER HAD ISSUES WITH PRIME/REWIND, AND THE DRIVE SUPPORT CAP WAS PROTRUDED. ADVISED THE CALLER TO DISCONTINUE THE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
126665 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-522CAL

Patients

Seq Age Sex Outcome Treatment
1