CUSTOM PAK
Report
- Report Number
- 1644019-2013-00027
- Event Type
- Malfunction
- Date Received
- March 20, 2013
- Date of Event
- February 1, 2013
- Report Date
- March 4, 2013
- Manufacturer
- ALCON - HOUSTON
- Product Code
- KYG
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
NO SAMPLE WERE RETURNED FOR EVALUATION; THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO THE MFR'S ACCEPTANCE CRITERIA. BECAUSE A SAMPLE WAS NOT RETUNED AND NO EVIDENCE OF NONCONFORMITY COULD BE FOUND IN THE LOT RECORD REVIEW, THE ROOT CAUSE FOR THE DULL KNIFE EXPERIENCED BY THE CUSTOMER CAN NOT BE DETERMINED. SOME POTENTIAL CAUSES ARE REUSE, IMPROPER HANDLING, OR CONTACT WITH ANOTHER INSTRUMENT ON THE INSTRUMENT TRAY DURING PROCEDURE SETUP. ALL KNIVES ARE 100% INSPECTED BY TRAINED OPERATORS USING A MINIMUM OF 10X MAGNIFICATION DURING MFG. ANY DEFECTS SUCH AS DAMAGED TIPS AND CUTTING EDGES, ARE REMOVED FROM THE LOT AND SCRAPPED. PENETRATION TESTING IS PERFORMED AND MONITORED DURING THE FINISHING PROCESS TO ENSURE THE SHARPNESS OF THE PRODUCT. (B)(4).
A CUSTOMER REPORTED THAT DURING A PROCEDURE, THE KNIFE WAS NOT CUTTING. THE PROCEDURE WAS COMPLETED WITH A BACKUP KNIFE. THERE WAS NO PT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 116754 | CUSTOM PAK | CONVENIENCE KIT | KYG | ALCON - HOUSTON | CUSTOM PAK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | CLEARCUT SIDEPORT FULL HANDLE |