FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3022916 · Received March 20, 2013

Report

Report Number
2518422-2013-00428
Event Type
Malfunction
Date Received
March 20, 2013
Date of Event
February 20, 2013
Report Date
February 20, 2013
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE VENTILATOR WAS RETURNED TO THE MFR FOR EVAL AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE APPEARED TO HAVE BEEN DROPPED; THEREBY CAUSING THE REPORTED ISSUE. THE DEVICE'S AC INLET CONNECTOR WILL BE REPLACED TO ADDRESS THE ISSUE. DEVICE HAS BEEN EVALUATED BUT NOT REPAIRED, PENDING CUSTOMER APPROVAL OF THE ESTIMATE.

Description of Event or Problem · 1

THE MFR REC'D INFO ALLEGING A VENTILATOR'S AC INLET CONNECTOR WAS DAMAGED. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
116698 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054260

Patients

Seq Age Sex Outcome Treatment
1