THERAPY COOL FLEX, 1304-CF-7-0.5(2)2-L-TE4BE1EB
Report
- Report Number
- 2030404-2013-00017
- Event Type
- Injury
- Date Received
- March 22, 2013
- Date of Event
- February 25, 2013
- Report Date
- February 25, 2013
- Manufacturer
- ST. JUDE MEDICAL, INC. (AF-IRVINE)
- Product Code
- OAD
- PMA / PMN Number
- P060019
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
ONE THERAPY ABLATION COOL FLEX CATHETER WAS RECEIVED FOR COMPLAINT EVALUATION. THERE WERE NO VISIBLE ANOMALIES NOTED. FUNCTIONAL TESTING REVEALED THE CATHETER MET ALL SPECIFICATIONS. THE CATHETER TIP WAS INVESTIGATED UNDER THE MICROSCOPE, WHICH IDENTIFIED NO ANOMALIES. THE DEVICE ALSO MET SPECIFICATIONS PRIOR TO RELEASE FROM SJM MANUFACTURING FACILITIES AS SUPPORTED BY THE DEVICE HISTORY RECORD. THE CAUSE OF THE REPORTED STEAM POP REMAINS UNKNOWN. THE IFU WARNS THE USER THAT VASCULAR PERFORATION IS AN INHERENT RISK OF ANY ELECTRODE PLACEMENT.
DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE USING A THERAPY COOL FLEX ABLATION CATHETER, A STEAM POP FOLLOWED BY A CARDIAC TAMPONADE OCCURRED. THE PHYSICIAN WAS PERFORMING ABLATION ON THE ANTERIOR PORTION OF THE LEFT ATRIUM WHEN A STEAM POP WAS HEARD BY THE STAFF. THE PATIENT BECAME HYPOTENSIVE AND AN ECHOCARDIOGRAM REVEALED A CARDIAC TAMPONADE. THE PHYSICIAN ATTEMPTED A PERICARDIOCENTESIS, WHICH WAS INSUFFICIENT TO RESOLVE THE TAMPONADE; THEREFORE, OPEN HEART SURGERY WAS PERFORMED TO SUCCESSFULLY TREAT THE TAMPONADE. THE PATIENT WAS STABILIZED AND IN GOOD CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 119503 | THERAPY COOL FLEX, 1304-CF-7-0.5(2)2-L-TE4BE1EB | CARDIAC ABLATION PERCUTANEOUS CATHETER | OAD | ST. JUDE MEDICAL, INC. (AF-IRVINE) | A088016 | 38794591 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Required Intervention | BRK NEEDLE: MODEL 407200, LOT 3885847| SWARTZ SL0 INTRODUCER| INQUIRY AFOCUS II: MODEL 81595, LOT 38595371| INQUIRY STEERABLE: MODEL 81104, LOT 3893050| COOL POINT PUMP| IBI T11 GENERATOR |