FDA Adverse Event
Malfunction
Summary report: N
ULTRASLIDE
MDR report key: 3022671
·
Received March 25, 2013
Report
- Report Number
- 3022671
- Event Type
- Malfunction
- Date Received
- March 25, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 25, 2013
- Manufacturer
- SKYTRON
- Product Code
- GDC
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- MI, US
Narratives
Description of Event or Problem · 1
WHILE WORKING IN A REVERSE TRENDELENBURG POSITION, THE BED SLID TOWARD THE FOOT. CLINICAL ENGINEERING THINKS THIS IS AN EQUIPMENT FAILURE. THE SLIDE CYLINDER SHAFT SNAPPED (BROKE). CLINICAL ENGINEERING THINKS THIS MAY BE DUE TO METAL FATIGUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 121369 | ULTRASLIDE | TABLE | GDC | SKYTRON | 3600B | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |