FDA Adverse Event Malfunction Summary report: N

SECURE II MED/SURG BED

MDR report key: 3022373 · Received March 27, 2013

Report

Report Number
0001831750-2013-02551
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
STRYKER MEDICAL-KALAMAZOO
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SUPPLEMENTAL SUBMITTED AS INVESTIGATION FOUND THAT THE FOWLER WAS ABLE TO BE LOWERED MANUALLY VIA THE CPR RELEASE, THEREFORE THE FOWLER WAS ABLE TO REACH ITS LOW, FLAT HEIGHT, WHICH IS THE PREFERRED POSITION FOR EMERGENCY MEDICAL INTERVENTION. THIS ISSUE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO SERIOUS INJURY OR DEATH IF IT WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE FOWLER WOULD NOT LOWER ELECTRICALLY DUE TO A MOTOR MALFUNCTION. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE FOWLER WOULD NOT LOWER DUE TO A MOTOR MALFUNCTION. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
125681 SECURE II MED/SURG BED BED, AC-POWERED ADJUSTABLE HOSPITAL FNL STRYKER MEDICAL-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1