FDA Adverse Event Injury Summary report: N

11MM/130 DEG TI CANN TROCH FIXATION NAIL 320MM/LEFT-STER

MDR report key: 3022257 · Received March 26, 2013

Report

Report Number
1719045-2013-00709
Event Type
Injury
Date Received
March 26, 2013
Date of Event
February 28, 2013
Report Date
February 28, 2013
Manufacturer
SYNTHES MONUMENT
Product Code
HSB
PMA / PMN Number
K011857
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. A REVIEW OF SYNTHES DEVICE HISTORY RECORDS FOR MANUFACTURING REVEALED NO COMPLAINT RELATED ISSUES.

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THE COMPLAINT CONDITION COULD NOT BE REPLICATED DURING THIS EVALUATION. THE RETURNED INSERTION HANDLE, AIMING ARM, BLADE GUIDE SLEEVE, TFN, AND HELICAL BLADE INSERTER WERE ASSEMBLED WITH KNOWN GOOD PARTS OF THE CONSTRUCT AND ENABLED AN ADEQUATE CONSTRUCT DURING THIS EVALUATION. THE HELICAL BLADE ALIGNED WITH AND PASSED THROUGH THE TFN HOLE AS INTENDED. THE DESIGN IS ADEQUATE FOR ITS INTENDED USE AND DID NOT CONTRIBUTE TO THIS COMPLAINT CONDITION.

Description of Event or Problem · 1

IT IS REPORTED THAT DURING A TFN PROCEDURE ON (B)(6) 2013, AS SURGEON WAS INSERTING THE HELICAL BLADE INTO THE NAIL, THE BLADE STOPPED. X-RAYS SHOWED THAT THE BLADE WAS HITTING THE WRONG PORTION OF THE NAIL. THE BLADE WOULD NOT GO INTO THE CENTER HOLE OF THE NAIL. ALSO, WHEN SURGEON ATTEMPTED TO REMOVE THE NAIL, THE NAIL WOULD NOT DISENGAGE FROM THE AIMING ARM. AT THAT POINT, THE SURGEON REMOVED ALL HARDWARE AND STARTED OVER WHAT A NEW SET OF INSTRUMENTS AND IMPLANTS. THIS ISSUE CAUSED SURGERY TO EXTEND AN ADDITIONAL 20 MINUTES. THIS IS 3 OF 8 REPORTS FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124441 11MM/130 DEG TI CANN TROCH FIXATION NAIL 320MM/LEFT-STER HSB SYNTHES MONUMENT 5014336

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention