PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-92115
- Event Type
- Death
- Date Received
- March 26, 2013
- Date of Event
- March 22, 2012
- Report Date
- August 2, 2021
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
THE INSULIN PUMP WAS RETURNED WITH NO BATTERY INSTALLED AND MISSING ADDRESS/SERIAL LABEL. THE INSULIN PUMP HAD INTERMITTENT ACT BUTTON RESPONSE. THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. THE INSULIN PUMP WAS RETURNED WITH A SCRATCHED DISPLAY WINDOW AND CRACKED BATTERY TUBE THREADS.
IT WAS REPORTED THAT THE CUSTOMER PASSED AWAY. IT WAS STATED THAT HIS DAUGHTER BELIEVES HIS LAST BLOOD GLUCOSE READING WAS 498MG/DL. THE DAUGHTER SATED THAT MAYBE HIS HEART STOPPED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 124229 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | Death |