PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3022180 · Received March 26, 2013

Report

Report Number
3004209178-2013-92115
Event Type
Death
Date Received
March 26, 2013
Date of Event
March 22, 2012
Report Date
August 2, 2021
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RETURNED WITH NO BATTERY INSTALLED AND MISSING ADDRESS/SERIAL LABEL. THE INSULIN PUMP HAD INTERMITTENT ACT BUTTON RESPONSE. THE INSULIN PUMP PASSED THE REWIND, BASIC OCCLUSION TEST, OCCLUSION TEST, PRIME TEST, EXCESSIVE NO DELIVERY TEST AND DISPLACEMENT TEST. THE INSULIN PUMP WAS RETURNED WITH A SCRATCHED DISPLAY WINDOW AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER PASSED AWAY. IT WAS STATED THAT HIS DAUGHTER BELIEVES HIS LAST BLOOD GLUCOSE READING WAS 498MG/DL. THE DAUGHTER SATED THAT MAYBE HIS HEART STOPPED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124229 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 53 YR Death