PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-92098
- Event Type
- Injury
- Date Received
- March 26, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 13, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NURSE
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
IT WAS REPORTED THAT THE CUSTOMER HAS ISSUES WITH THE INSULIN PUMP, AND HE WAS IN THE EMERGENCY ROOM DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 500MG/DL. TROUBLESHOOTING WAS PERFORMED. THE TIME, DATE, AND BOLUS WIZARD SETTINGS WERE CORRECT. REVIEWED THE ALARM HISTORY AND FOUND A NO DELIVERY ALARM. THE NURSE STATED THAT THE DRIVE SUPPORT CAP APPEARS NORMAL. ASSISTED THE CALLER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED THE HIGH PRESSURE TEST AND PASSED. INSTRUCTED THE NURSE TO REMOVE THE CANNULA AND IT WAS NOT BENT OR OCCLUDED. ADVISED THE NURSE TO CHANGE THE ENTIRE INFUSION SET AND RESERVOIR. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 124699 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR | Hospitalization |