HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE WIRE
Report
- Report Number
- 2024168-2013-01742
- Event Type
- Injury
- Date Received
- March 26, 2013
- Date of Event
- March 1, 2013
- Report Date
- March 1, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- DQX
- PMA / PMN Number
- K013833
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). IT WAS INDICATED THAT THE DEVICE WILL NOT BE RETURNING, THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. HOWEVER, A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES WITH THE REPORTED LOT AND A REVIEW OF THE COMPLAINT HISTORY DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT AN UNSPECIFIED 20-28 MM IN LENGTH BARE METAL STENT IMPLANTED IN THE PROXIMAL-MID RIGHT CORONARY ARTERY (RCA) AT AN UNSPECIFIED TIME WAS RESTENOSED. WITH A FEMORAL ARTERY ACCESS, A BALANCED MIDDLEWEIGHT (BMW) GUIDE WIRE WITH A NON-ABBOT SUPPORT CATHETER WAS ADVANCED INTO THE TOTALLY OCCLUDED PROXIMAL RCA THROUGH THE UNSPECIFIED BARE METAL STENT STRUTS AND MET RESISTANCE WITH THE UNSPECIFIED STENT STRUTS CAUSING THE BMW GUIDE WIRE TO ENTER THE SUBINTIMAL SPACE OF THE PATIENT. THERE WAS A FLUOROSCOPY DONE WITH DYE AND IT WAS DECIDED TO ABANDON THE PROCEDURE. THE BMW AND THE NON-ABBOT SUPPORT CATHETER WERE REMOVED AND UPON REMOVAL THE BMW GUIDE WIRE WAS FOUND SEPARATED PROXIMAL TO THE MARKER BAND. THE DISTAL PART OF THE BMW GUIDE WIRE WAS FOUND STILL IN THE SUBINTIMAL SPACE, NOT IN THE PATIENTS VASCULAR SYSTEM; THEREFORE, IT WAS DECIDED TO LEAVE THE SEPARATED PORTION OF THE GUIDE WIRE INSIDE THE PATIENTS ANATOMY. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 123497 | HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL GUIDE WIRE | GUIDE WIRE | DQX | AV-TEMECULA-CT | 2112871 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Other | OTHER: CROSS BOSSSTENT: UNSPECIFIED BARE METAL |