FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 3022094 · Received March 26, 2013

Report

Report Number
2955842-2013-00968
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
February 21, 2013
Report Date
February 25, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING CONFIRMED A CABLE SNAPPED. ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STUCK OUT AT THE INSTRUMENTS WRIST. THE OTHER CABLES AT THE WRIST WERE NOT DAMAGED. AN ADDITIONAL FINDING WERE SCRATCHES ON THE PULLEY. THERE WERE SOME SCRATCHES ON THE SURFACE OF THE DISTAL PULLEY. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, A CABLE SNAPPED AT DISTAL END OF THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124548 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10121128 241

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES