FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3022008 · Received March 26, 2013

Report

Report Number
2032227-2013-01145
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
March 14, 2013
Report Date
March 14, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST AND ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A LOOSE DRIVE SUPPORT DISK. NO ALARMS WERE NOTED. THE INSULIN PUMP HAD A SEVERELY SCRATCHED DISPLAY WINDOW AND CASE. THE INSULIN PUMP PASSED THE MOTOR TEST.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM FOLLOWED BY ANOTHER ALARM DURING THE PRIME AND REWIND PROCESS. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124673 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722NAS

Patients

Seq Age Sex Outcome Treatment
1 37 YR