FDA Adverse Event Injury Summary report: N

SETROX S 45

MDR report key: 3021936 · Received March 26, 2013

Report

Report Number
1028232-2013-00873
Event Type
Injury
Date Received
March 26, 2013
Date of Event
March 13, 2013
Report Date
March 14, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ASSOCIATED WITH THE MANUFACTURE OF THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

THERE WAS AN RA LEAD REVISION DONE ON (B)(6) 2013 DUE TO DISLODGMENT. THIS LEAD REMAINS ACTIVELY IN SERVICE. THERE WERE NO ADVERSE PATIENT EVENTS REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
123007 SETROX S 45 PACER LEAD NVN BIOTRONIK SE & CO. KG 350973

Patients

Seq Age Sex Outcome Treatment
1 51 YR Hospitalization