FDA Adverse Event Malfunction Summary report: N

LINOX SMART SD 65/16

MDR report key: 3021927 · Received March 26, 2013

Report

Report Number
1028232-2013-00846
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
December 12, 2012
Report Date
March 15, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVY
PMA / PMN Number
P980023
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PERFORMANCE OF THE LEAD UNDER COMPLAINT WAS SCRUTINIZED. THE ANALYSIS OF THE LEAD DEMONSTRATED A RUBBED THROUGH INSULATION AND A FRACTURED CONDUCTOR CABLE TO THE RV SHOCK COIL AT THIS POINT. THIS DAMAGE CAN BE CONSIDERED AS THE ROOT CAUSE FOR THE OBSERVED HIGH SHOCK IMPEDANCE MENTIONED IN THE COMPLAINT DESCRIPTION. BASED ON THE CHARACTERISTICS OF THE DAMAGE, IT IS REASONABLE TO ASSUME THAT THE LEAD HAD BEEN SUBJECT TO SEVERE MECHANICAL STRESS IN THE IMPLANTED STATE. INTERACTION BETWEEN THE LEAD BODY AND THE TRICUSPID VALVE SHOULD BE TAKEN INTO CONSIDERATION. DIAGNOSTIC IMAGES CLARIFYING THIS ASSUMPTION WERE NOT AVAILABLE FOR ANALYSIS. FURTHER DAMAGES OCCURRED MOST LIKELY DURING THE EXPLANTATION PROCEDURE. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

OUS MDR - ON (B)(6) 2012, DURING A FOLLOW-UP, THIS LEAD PRESENTED A HIGH SHOCK IMPEDANCE (> 150 OHMS) AND A HIGH PACE/SENSE IMPEDANCE ( > 3000 OHMS). DURING THE REPLACEMENT PROCEDURE, IT WAS NOTED THAT ABOVE THE DISTAL COIL THE INSULATION WAS DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
123964 LINOX SMART SD 65/16 ICD LEAD NVY BIOTRONIK SE & CO. KG 359066

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization