FDA Adverse Event Injury Summary report: N

7.0MM TI DUAL-OPENING SCREW 50MM THRD LENGTH F/6.0MM RODS

MDR report key: 3021856 · Received March 26, 2013

Report

Report Number
2530088-2013-00366
Event Type
Injury
Date Received
March 26, 2013
Date of Event
February 25, 2013
Report Date
February 25, 2013
Manufacturer
SYNTHES BRANDYWINE
Product Code
NKB
PMA / PMN Number
K082572
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE PRODUCT WENT TO AN ENGINEER TO BE EVALUATED: THE 388.612 IS CLEARLY BENT AT THE INTERNAL THREADED SHAFT. ATTEMPTING TO APPLY A NUT TO THE CONSTRUCT OVER THIS SHAFT COULD POSSIBLY LEAD TO CROSS-THREADING OF THE NUT ON THE SCREW AS THE AXIS IS NO LONGER ALIGNED. IT IS NOT CLEAR HOW OR WHEN THE SHAFT WAS BENT, BUT GIVEN THE MIS-ALIGNMENT, CROSS-THREADING COULD OCCUR WHICH WOULD BE CONDUCIVE TO THE DIFFICULTY EXPERIENCED BY THE SURGEON WHEN ATTEMPTING TO PLACE THE NUT. THE MATERIALS WERE REVIEWED AND ARE ADEQUATE FOR THE DEVICES INTENDED USE.

Additional Manufacturer Narrative · 1

DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. DAMAGE ON THE M8 THREAD, SCRATCHES AND NICKS ON TOP OF THE SCREW AND UPPER BODY DIAMETER. ALSO THERE IS ANODIZE FINISH MISSING ON THE DAMAGED THREAD AREA AND ON THE EDGES OF THE SIDE OPENING. BECAUSE OF THE RETURNED CONDITION OF THE SCREW, NOT ALL RELEVANT FEATURES CAN BE VERIFIED.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DHR REVIEW FOUND NO RELEVANT ISSUES THAT WOULD RESULT IN THIS PRODUCT COMPLAINT. OTHER PRODUCT CODES: MNI, MNH, KWP AND KWQ.

Description of Event or Problem · 1

THIS IS 10 OF 12 REPORTS FOR COMPLAINT (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON WAS PERFORMING L2-L4 POSTERIOR FUSION WITH USS DUAL OPENING. WHEN CONNECTING THE RODS, HE COULD NOT GET THE NUT TO ENGAGE ON THE RIGHT L4 7.0 X 50MM DO SCREW. THE SURGEON ATTEMPTED TO ENGAGE FIVE USS DO NUTS BUT THE THREADS HAD STRIPPED AT THE TOP OF THE SCREW. THE SURGEON REPLACED THE 7.0 X 50MM SCREW WITH A 8.0 X 50MM DO SCREW AND ATTEMPTED 2 MORE SETS OF NUTS AND COLLARS BUT THE NUTS WOULD STILL NOT ENGAGE THE THREADS ON THE SCREW. THE SURGEON STRIPPED THE THREADS ON THE 8.0 X 50MM SCREW AND REPLACED IT WITH ANOTHER 8.0 X 50MM SCREW AFTER THE BENDING OF THE ROD IN SITU. WHEN THE SURGEON REMOVED THE SCREW IT WAS NOTICED THAT THE GUIDE STICK WAS BENT, AND STILL REMAINS ON THE SCREW. THE PROCEDURE WAS REPORTEDLY COMPLETED WITH NO HARM TO PATIENT. THIS IS 10 OF 12 REPORTS FOR THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
123756 7.0MM TI DUAL-OPENING SCREW 50MM THRD LENGTH F/6.0MM RODS NKB SYNTHES BRANDYWINE 6913769

Patients

Seq Age Sex Outcome Treatment
1 65 YR Required Intervention