FDA Adverse Event Malfunction Summary report: N

DEXTRUS 4137

MDR report key: 3021808 · Received March 26, 2013

Report

Report Number
1028232-2013-00742
Event Type
Malfunction
Date Received
March 26, 2013
Date of Event
January 27, 2013
Report Date
March 11, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ANALYSIS OF THE LEAD DEMONSTRATED A DAMAGED INSULATION AND A FRACTURED OUTER COIL AT A DISTANCE OF SOME 27.5 CM DISTAL TO THE IS-1 CONNECTOR PIN. IN THAT SECTION, THE LEAD BODY WAS FOUND SQUEEZED AND DEFORMED. BASED ON THE CHARACTERISTICS AS WELL AS THE LOCATION OF THESE DAMAGES, IT IS REASONABLE TO ASSUME THAT THE LEAD HAD BEEN SUBJECT TO EXCESSIVE MECHANICAL FORCES AS THE RESULT OF CLAVICULAR FIRST RIB ENTRAPMENT. FURTHER INSPECTION REVEALED INSULATION DAMAGES IN THE DISTAL PART OF THE LEAD, WHICH WAS MOST LIKELY CAUSED BY AN INTERACTION BETWEEN THE LEAD AND THE TRICUSPID VALVE. THESE DAMAGES CAN BE CONSIDERED AS THE ROOT CAUSE OF THE OBSERVED ISSUES AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE ANALYSIS DID NOT REVEAL ANY SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS RIGHT VENTRICULAR LEAD PRESENTED IN THE ER FOR AN UNSPECIFIED REASON. UPON REVIEW, THE RV LEAD DISPLAYED NOISE AS WELL AS PACING IMPEDANCE MEASUREMENTS OF 160 OHMS. NO PACING INHIBITION WAS OBSERVED, AS THE PATIENTS OWN INTRINSIC RHYTHM WAS APPROPRIATE. A REVISION PROCEDURE WAS PERFORMED. THE LEAD WAS EXPLANTED. UPON VISUAL INSPECTION, THE PHYSICIAN OBSERVED EROSION ON THE LEAD INSULATION. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124782 DEXTRUS 4137 PACER LEAD NVN BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization